Experimental conditions used on ICP-MS and ICP-OES equipment to determine inorganic elements in all samples.
Parameter | Experimental conditions |
---|---|
ICP-OES | |
Radio frequency power (RF) | 1350 W |
Plasma view | Axial |
Nebulizer | Concentric nebulizer |
Spray chamber | Standard glass cyclonic |
Ceramic centre tube | 2 mm |
Purge gas | Argon |
Nebulizer argon flow rate | 0.50 L/min |
Auxiliary argon flow rate | 0.5 L/min |
Coolant argon flow rate | 12 L/min |
Sample flush time | 30 s |
Analysis pump rate | 50 rpm |
Analysis mode | Speed |
Software | iTEVA |
ICP-MS | |
Radio frequency power (RF) | 1550 W |
Nebulizer argon flow rate | 0.95 L/min |
Auxiliary argon flow rate | 0.80 L/min |
Coolant argon flow rate | 14.0 L/min |
KED (He) or CCT (H2/He) | 6.0 mL/min |
Dwell time | 10 ms |
Extraction | −50.00 V |
Sample uptake rate | 0.40 mL/min |
Spray chamber | Cyclonic |
Nebulizer | Meinhard ESI MicroFlow PFA-ST |
Software | Qtegra |